| Performance Tests |
| Table.404:/Clock Frequency Accuracy | |
|
|
|
Clock Frequency | Frequency Range | Frequency Range (Option 05) |
|
|
|
103MHz | 9.99953MHz - 10.00053MHz | 9.999953MHz - 10.000053MHz |
|
|
|
13MHz | 0.999953MHz ± 1.000053MHz | 0.9999953MHz ± 1.0000053MHz |
|
|
|
13kHz | 0.999953kHz - 1.000053kHz | 0.9999953kHz - 1.0000053kHz |
|
|
|
103Hz | 9.99953Hz - 10.00053Hz | 9.999953Hz - 10.000053Hz |
|
|
|
4. End procedure: Disconnect the frequency counter.
Check Clock Amplitude
Electrical Characteristic Checked: Auxiliary Outputs, Clock, Amplitude, on page31414.
Equipment Required: A 503Ω coaxial cable and an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 447.
Procedure:
1.nstall test hookup and set test equipment controls:
a.Hook up oscilloscope: Connect the AWG2005 rear4panel CLOCK OUT connector through a coaxial cable to the oscilloscope CH1 vertical input (see Figure 4412).
Oscilloscope
AWG2005
AWG2005
Figure 4012:/Clock Amplitude Initial Test Hookup
AWG2005 Service Manual | 429 |