Performance Tests

Vertical

CH1

CH1 Coupling

DC

CH1 Scale

500:mV/div.

CH1 Input Impedance

50:Ω

Horizontal

 

Sweep

50:μs/div.

Trigger

 

Source

CH1

Coupling

DC

Slope

Positive

Level

500:mV

Mode

Auto

2.Set the AWG2005 controls and select the waveform file:

a.Initialize AWG2005 controls: Push UTILITY￿Misc￿Config...￿Re) set to Factory￿O.K.

b.Select waveform file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Turn the general purpose knob to highlight the MRK.WFM file.

￿Push ENTER to select the file.

3.Check rear)panel CH1 MARKER OUT amplitude:

a.Check CH1 MARKER OUT pulse amplitude:

￿Check that the pulse amplitude of the displayed waveform is 2:Vp;p or greater.

4.Check rear)panel CH2 MARKER OUT pulse amplitude:

a.Check CH2 MARKER OUT pulse amplitude:

￿Move the coaxial cable from the AWG2005 rear;panel CH1 MARKER OUT connector to the rear;panel CH2 MARKER OUT connector.

￿Check that the pulse amplitude of the displayed waveform is 2:V or greater.

5.Check Option 02: If the AWG2005 has CH3 and CH4, repeat this entire test, selecting the AWG2005 waveform and setting controls for CH3 and CH4 and checking:

￿ Rear;panel CH3 MARKER pulse amplitude

￿ Rear;panel CH4 MARKER pulse amplitude

6.End procedure: Disconnect the oscilloscope.

AWG2005 Service Manual

4￿37