Performance Tests
Vertical | CH1 |
CH1 Coupling | DC |
CH1 Scale | 500:mV/div. |
CH1 Input Impedance | 50:Ω |
Horizontal |
|
Sweep | 50:μs/div. |
Trigger |
|
Source | CH1 |
Coupling | DC |
Slope | Positive |
Level | 500:mV |
Mode | Auto |
2.Set the AWG2005 controls and select the waveform file:
a.Initialize AWG2005 controls: Push UTILITYMiscConfig...Re) set to FactoryO.K.
b.Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the MRK.WFM file.
Push ENTER to select the file.
3.Check rear)panel CH1 MARKER OUT amplitude:
a.Check CH1 MARKER OUT pulse amplitude:
Check that the pulse amplitude of the displayed waveform is 2:Vp;p or greater.
4.Check rear)panel CH2 MARKER OUT pulse amplitude:
a.Check CH2 MARKER OUT pulse amplitude:
Move the coaxial cable from the AWG2005 rear;panel CH1 MARKER OUT connector to the rear;panel CH2 MARKER OUT connector.
Check that the pulse amplitude of the displayed waveform is 2:V or greater.
5.Check Option 02: If the AWG2005 has CH3 and CH4, repeat this entire test, selecting the AWG2005 waveform and setting controls for CH3 and CH4 and checking:
Rear;panel CH3 MARKER pulse amplitude
Rear;panel CH4 MARKER pulse amplitude
6.End procedure: Disconnect the oscilloscope.