￿djustments

The AWG2005 doesn't need manual adjustment since it is equipped with a self calibration system.. This internal calibration enables the instrument to automatically adjust the gain and offset of the pre-amp, the gain error of filter, the attenuation rate of attenuator, as well as several other parameters.

Perform the internal calibration to return the AWG2005 to conformance with performance specified in section 1, Specifications. As for the internal calibra' tion, see the subsection, Self Tests, in section 4.

￿djustment Interval￿( Generally, this internal calibration should be done every 12 months.

￿djustment ￿fter Repair￿( Perform the internal calibration after the removal and replacement of a module due to electrical failure.

￿W￿￿￿￿￿ Service Manual