Figure 428:1Autostep Mode Initial Test Hookup
b S : |
|
Vertical | CH1 |
CH1 coupling | DC |
CH1 scale | 05/V/div |
CH1 input impedance | 50/Ω |
Horizontal |
|
Sweep | 50/μs/div |
Trigger |
|
Source | CH2 |
Coupling | DC |
Slope | Positive |
Level | 100/mV |
Mode | Auto |
2 S AWG2005 w :
a Iz AWG2005 : Push UTILITYMiscConfig...Re2 set to FactoryO.K.
b My AWG2005 :
Push MODEAutostepSlope to highlight Positive
Push Run from the side menu to highlight Continuous
Select Level from the side menu and turn the general purpose knob to select 1/V
Select Select Autostep File from the side menu
Turn the general purpose knob to highlight the MODE_AST.AST file
Push ENTER to select the file
3 Ck : Push the AWG2005 MANUAL TRIGGER button and check that the oscilloscope momentarily displays a sine wave with a different frequency and amplitude each time you push the button
4 E : Disconnect the oscilloscope