Inspection and Cleaning
CAUTION
To prevent damaging the plastics used in the AWG2005 do not use chemical cleaning agents. Use only deionized water when cleaning the menu buttons or front'panel buttons. Use a ethyl alcohol solu' tion as a cleaner, and rinse with deionized water.
To prevent damaging AWG2005 components, do not use high'pres' sure compressed air when cleaning dust from the interior of the AWG2005. (High pressure air can cause electrostatic discharge.) Instead, use low pressure compressed air (about 9 psi).
Inspection#Exterior
Inspect the outside of the AWG for damage wear and missing parts using Table 6* as a guide If the AWG appears to have been dropped or otherwise abused check it thoroughly to verify correct operation and performance Repair any defects that may cause personal injury or lead to further damage to the AWG
Table 6"2:!External Inspection Check List
Item | Inspect For | Repair ction |
|
|
|
Cabinet front panel and | Cracks scratches deformations dam* | Replace defective module |
cover | aged hardware or gaskets |
|
|
|
|
Front*panel knobs | Missing damaged or loose knobs | Repair or replace missing or defective |
|
| knobs |
|
|
|
Connectors | Broken shells cracked insulation and | Replace defective modules Clear or |
| deformed contacts Dirt in connectors | wash out dirt |
|
|
|
Carrying handle cabinet | Correct operation | Replace defective module |
feet |
|
|
|
|
|
Accessories | Missing items or parts of items bent | Replace damaged or missing items |
| pins broken or frayed cables and | frayed cables and defective modules |
| damaged connectors |
|
|
|
|
66 | Maintenance |