Inspection and Cleaning

CAUTION

To prevent damaging the plastics used in the AWG2005 do not use chemical cleaning agents. Use only deionized water when cleaning the menu buttons or front'panel buttons. Use a ethyl alcohol solu' tion as a cleaner, and rinse with deionized water.

To prevent damaging AWG2005 components, do not use high'pres' sure compressed air when cleaning dust from the interior of the AWG2005. (High pressure air can cause electrostatic discharge.) Instead, use low pressure compressed air (about 9 psi).

Inspection￿#￿Exterior

Inspect the outside of the AWG￿￿￿￿ for damage￿ wear￿ and missing parts￿ using Table 6*￿ as a guide￿ If the AWG￿￿￿￿ appears to have been dropped or otherwise abused￿ check it thoroughly to verify correct operation and performance￿ Repair any defects that may cause personal injury or lead to further damage to the AWG￿￿￿￿￿

Table 6"2:!External Inspection Check List

Item

Inspect For

Repair ￿ction

 

 

 

Cabinet￿ front panel￿ and

Cracks￿ scratches￿ deformations￿ dam*

Replace defective module￿

cover

aged hardware or gaskets￿

 

 

 

 

Front*panel knobs

Missing￿ damaged￿ or loose knobs￿

Repair or replace missing or defective

 

 

knobs￿

 

 

 

Connectors

Broken shells￿ cracked insulation￿ and

Replace defective modules￿ Clear or

 

deformed contacts￿ Dirt in connectors￿

wash out dirt￿

 

 

 

Carrying handle￿ cabinet￿

Correct operation￿

Replace defective module￿

feet

 

 

 

 

 

Accessories

Missing items or parts of items￿ bent

Replace damaged or missing items￿

 

pins￿ broken or frayed cables￿ and

frayed cables￿ and defective modules￿

 

damaged connectors￿

 

 

 

 

6￿6

Maintenance