Instructions for Operation

Before servicing the AWG2005, read the following operating instructions.

These instructions are at the level appropriate for servicing the AWG2005.

The user manual contains complete operator instructions.

In addition, section 4, Performance Verification, includes instructions for making the front8panel settings required to check AWG2005 characteristics.

How to Power On

To power8on the AWG2005, follow these steps:

 

1. Set the PRINCIPAL POWER SWITCH (on the back of the AWG2005) to

 

the ON position. This switch is the main power switch; it routes power to

 

the standby circuit in the AWG2005.

 

2. Then, press the ON/STBY (standby) switch on the front (lower8left cor8

 

ner) of the AWG2005. This switch applies power to the remaining circuits

 

of the AWG2005. Allow at least 20 minutes for the AWG2005 to warm up.

 

 

 

 

 

 

 

 

 

WARNING

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

To avoid personal shock hazard, turn off both the ON/STBY switch

 

and the PRINCIPAL POWER SWITCH before servicing. The PRINCI2

 

PAL POWER SWITCH on the rear panel is the true power discon2

 

nect switch. The ON/STBY (standby) switch simply toggles

 

operation on and off. When connected to a power source and when

 

the PRINCIPAL POWER SWITCH is on, the internal power supplies

 

and much of the other circuitry of the AWG2005 remain energized

 

regardless of the setting of the ON/STBY switch.

 

To avoid personal shock hazard, set the PRINCIPAL POWER

 

SWITCH off before connecting or disconnecting the line cord to or

 

from the power source.

Internal Diagnostics

At power8on, the AWG2005 performs internal start8up diagnostics. These

Routines

diagnostics check internal circuit function and report any failures. In addi8

tion, you can initiate internal diagnostics; these diagnostics differ from the

 

 

start8up diagnostics in that they do more extensive memory checking.

 

The AWG2005 also contains internal calibration routines, which check

 

internal circuit function and adjust calibration constants. Run these calibra8

 

tion routines whenever the AWG2005 undergoes a temperature change. For

￿WG￿￿￿￿ Service Manual