Self Tests

Figure 4￿1:￿Diagnostics Menu

b.Run the diagnostics: Select Execute from the side menu. This executes all the AWG2005 diagnostics automatically.

c.Wait: The internal diagnostics do an extensive verification of AWG2005 functions. While this verification progresses, the screen displays the clock icon. When finished, the resulting status appears on the screen.

d.Confirm that no failures are found: Verify that no failures are found and reported on0screen. If the diagnostics displays FAIL as the result of any test, use the instructions in section 6, Troubleshooting, to identify the faulty module. If the diagnostics display an error code, contact your nearest representative.

2.Return to regular service: Push a button (other than UTILITY) in the MENU column to exit the diagnostic menu.

4￿4

Performance Verification