Performance Tests
Check Triggered Mode
Electrical Characteristic Checked: Operating modes, Triggered, on
page6175.
Equipment Required: Two 506Ω coaxial cables, a function generator, and an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 477.
Procedure:
1.Install the test hookup and set test equipment controls:
a.Hook up the oscilloscope: Connect the AWG2005 CH1 output connector through the coaxial cable to the CH1 vertical input con7 nector on the oscilloscope.
b.Hook up the function generator:
Connect the AWG2005 TRIGGER INPUT connector though a coaxial cable to the function generator output connector (see Figure 474).
Function Generator
FG 5010
AWG2005 | Oscilloscope |
AWG2005
Figure 4"4:!Triggered Mode Initial Test Hookup
c.Set the oscilloscope controls:
Vertical: | CH1 |
CH1 coupling: | DC |
CH1 scale | 0.26V/div. |
CH1 input impedance: | 506Ω |
Horizontal |
|
Sweep | 506μs/div. |