Performance Tests

Check Triggered Mode

Electrical Characteristic Checked: Operating modes, Triggered, on

page6175.

Equipment Required: Two 506Ω coaxial cables, a function generator, and an oscilloscope.

Prerequisites: The AWG2005 meets the prerequisites listed on page 477.

Procedure:

1.Install the test hookup and set test equipment controls:

a.Hook up the oscilloscope: Connect the AWG2005 CH1 output connector through the coaxial cable to the CH1 vertical input con7 nector on the oscilloscope.

b.Hook up the function generator:

￿Connect the AWG2005 TRIGGER INPUT connector though a coaxial cable to the function generator output connector (see Figure 474).

Function Generator

FG 5010

AWG2005

Oscilloscope

AWG2005

Figure 4"4:!Triggered Mode Initial Test Hookup

c.Set the oscilloscope controls:

Vertical:

CH1

CH1 coupling:

DC

CH1 scale

0.26V/div.

CH1 input impedance:

506Ω

Horizontal

 

Sweep

506μs/div.

4￿14

Performance Verification