Performance Tests

Procedure:

1.Install test hookup and set test equipment controls:

a. Hook up the oscilloscope: Connect the AWG2005 CH1 output connector through the coaxial cable to the CH1 vertical input con4 nector on the oscilloscope (see Figure 447).

AWG2005

Oscilloscope

 

 

 

AWG2005

Figure 407:/Waveform Advance Mode Initial Test Hookup

b. Set oscilloscope controls:

 

Vertical

CH1

CH1 coupling

DC

CH1

0.23V/div.

CH1 input impedance

503Ω

Horizontal

 

Sweep

503μs/div.

Trigger

 

Source

CH1

Coupling

DC

Slope

Positive

Level

03V

Mode

Auto

2.Set the AWG2005 controls and select the waveform file:

a.Initialize AWG2005 controls: Push UTILITY￿Misc￿Config...￿Re0 set to Factory￿O.K.

b.Set AWG2005 controls:

￿Push MODE￿Waveform Advance￿Slope to highlight Posi0 tive.

￿Select Level from the side menu, and turn the general purpose knob to select a 1.03V level.

￿Check that the side menu Run highlights Continuous. If neces4 sary, push Run to select Continuous.

c.Select waveform file:

AWG2005 Service Manual

4￿19