Performance Tests

3.Enable the function generator output: Turn on the function generator output.

4.Check external AM operation:

￿Check that the step number displayed on the AWG2005 MODE menu is Step: 1 (see Figure 4510). If it is not, push MANUAL TRIG% GER to step though the autostep file steps until Step 1 is displayed.

Step Display

Figure 4%1￿:$MODE Menu Autostep Setting

￿Check that the DMM reading is in the range from 2.375 to 2.6254V (100% modulation).

￿Set the function generator offset value to 04V. Check that the DMM reading is in the range from 1.125 to 1.3754V (50% modulation).

￿Set the function generator offset value to -14V. Check that the DMM voltage reading is in the range from -0.125 to 0.1254V (0% modula5 tion).

5.End procedure: Keep the test connections and instrument settings for the next check.

4￿24

Performance Verification