Performance Tests
6.End procedure: Turn off the function generator output, and disconnect the function generator and oscilloscope.
Check Gated Mode
Electrical Characteristic Checked: Operating modes, Gated, on
page3145.
Equipment Required: Three 503Ω coaxial cables, a 50 Ω precision ter4 mination, a function generator, and an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 447.
Procedure:
1.nstall test hookup and set test equipment controls:
a.Hook up the oscilloscope: Connect the AWG2005 CH1 output connector through the coaxial cable to the CH1 vertical input con4 nector on the oscilloscope.
b.Hook up the function generator: Connect the function generator output to both the AWG2005 TRIGGER INPUT and the oscilloscope CH2 input through a coaxial cable and a dual input coupler (see Figure 445).
AWG2005 | Oscilloscope | ||
|
|
|
|
|
|
|
|
AWG2005
Function Generator
FG 5010
Figure 4$5:#Gated Mode Initial Test Hookup
c. Set oscilloscope controls: |
|
Vertical | CH1 |
CH1 coupling | DC |
CH1 scale | 0.53V/div. |
CH1 input impedance | 503Ω |
416 | Performance Verification |