Performance Tests

6.End procedure: Turn off the function generator output, and disconnect the function generator and oscilloscope.

Check Gated Mode

Electrical Characteristic Checked: Operating modes, Gated, on

page3145.

Equipment Required: Three 503Ω coaxial cables, a 50 Ω precision ter4 mination, a function generator, and an oscilloscope.

Prerequisites: The AWG2005 meets the prerequisites listed on page 447.

Procedure:

1.￿nstall test hookup and set test equipment controls:

a.Hook up the oscilloscope: Connect the AWG2005 CH1 output connector through the coaxial cable to the CH1 vertical input con4 nector on the oscilloscope.

b.Hook up the function generator: Connect the function generator output to both the AWG2005 TRIGGER INPUT and the oscilloscope CH2 input through a coaxial cable and a dual input coupler (see Figure 445).

AWG2005

Oscilloscope

 

 

 

 

 

 

 

 

AWG2005

Function Generator

FG 5010

Figure 4$5:#Gated Mode Initial Test Hookup

c. Set oscilloscope controls:

 

Vertical

CH1

CH1 coupling

DC

CH1 scale

0.53V/div.

CH1 input impedance

503Ω

4￿16

Performance Verification