Performance Tests
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Highlight the MODE_DV.SEQ file using the general purpose knob.
Push ENTER to select the file.
3.Tur te AWG2005 CH1 utut: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Ceck wavefr advace: Repeatedly push the AWG2005 MNUL TRIGGER button, and check that the oscilloscope displays a continuous sine wave that switches between two frequencies at each manual trig8 ger.
5.Ed rcedure: Disconnect the oscilloscope.
Check utostep Mode
Electrical Characteristic Checked: Operating mode, Autostep, on
page7185.
Equipment Required: Two 507Ω coaxial cables and an oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 487.
Procedure:
1.Ista test ku ad set test equet ctrs: a. Hk u te scsce:
Connect the AWG2005 CH1 output through a coaxial cable to the oscilloscope CH1 vertical input.
Connect the AWG2005 rear8panel CH1 MARKER OUT output through a coaxial cable to the oscilloscope CH2 vertical input (see Figure 488). The CH1 MARKER OUT signal will serve as an external trigger signal for the oscilloscope.
420 | Performance Verification |