Performance Tests
Trigger |
|
Source | CH1 |
Coupling | DC |
Slope | Positive |
Level | |
Mode | Auto |
d.Set the function generator controls:
Function | Square |
Mode | Continuous |
Parameter |
|
Frequency | 16kHz |
Amplitude | 46V |
Offset | 26V |
Output | Off |
2.Set AWG2005 controls and select the waveform file:
a.Initialize AWG2005 controls: Push UTILITYMiscConfig...Re. set to FactoryO.K.
b.Modify the AWG2005 default settings:
Push MODETriggeredSlope to select Positive slope.
Select Level from the side menu and turn the general purpose knob to select a 16V trigger level.
c.Select the file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Highlight the MODE.WFM file using the general purpose knob.
Push ENTER to select the file.
3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check triggered mode with manual triggering: Push the AWG2005 MANUAL TRIGGER button and check that when the button is pushed, the oscilloscope displays a one7cycle sine wave.
5.Check triggered mode with external triggering:
a.Change the oscilloscope trigger mode to Normal.
b.Enable function generator output: Turn on the function generator output.
c.Check triggering: Check that for each trigger supplied by the func7 tion generator, the oscilloscope displays a one7cycle sine wave.
AWG2005 Service Manual | 415 |