Performance Tests

Function Generator

FG 5￿1￿

AWG2005

Oscilloscope

AWG2￿￿5

Figure 4.18:-External Trigger Level Accuracy Initial Test Hookup

c. Set oscilloscope controls:

 

Vertical

CH1

CH1 Coupling

DC

CH1 Scale

￿.20V/div.

CH1 Input Impedance

5￿0Ω

Horizontal

 

Sweep

5￿0μs/div.

Trigger

 

Source

CH1

Coupling

DC

Slope

Positive

Level

￿0V

Mode

Auto

d.Set function generator controls:

Function

Square

Mode

Continuous

Parameter

 

Frequency

10kHz

Amplitude

￿0V

Offset

￿.60V

Output

Off

2.Select the AWG2005 waveform file and set AWG2005 controls:

a.Initialize AWG2005 controls: Push UTILITY￿Misc￿Config...￿Re. set to Factory￿O.K.

b.Modify AWG2005 default settings:

￿ Push MODE￿Gated￿Polarity to highlight Positive.

4￿40