Performance Tests
Select Level from the side menu, and turn the general purpose knob to select 19V. (You can also use the numeric and units keys to select 19V; then push ENTER.)
c.Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the TRG_IN.WFM file.
Push ENTER to select the file.
3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check external trigger high level:
a.Adjust oscilloscope controls: Press and hold the AWG2005 MANU+ AL TRIGGER button and adjust the oscilloscope vertical and hori: zontal position to display the waveform from the AWG2005. Release the MANUAL TRIGGER button.
b.Enable function generator output: Turn on the function generator output.
c.Check external trigger level accuracy:
Gradually increment the function generator offset level until a waveform is displayed on the oscilloscope.
Check that that the function generator offset level is from 0.85 to 1.159V when the waveform is first displayed.
5.Check external trigger low level:
a.Change the function generator controls:
Parameter
Offset-0.69V
b.Change the AWG2005 controls:
Push MODEPolarity to highlight Negative.
Select Level from the side menu, and turn the general purpose knob to select
c.Check external trigger level accuracy:
Gradually decrease the function generator offset level until a waveform is displayed on the oscilloscope.
Check that that the function generator offset level is from
AWG2005 Service Manual | 441 |