Performance Tests

￿Select Level from the side menu, and turn the general purpose knob to select 19V. (You can also use the numeric and units keys to select 19V; then push ENTER.)

c.Select waveform file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Turn the general purpose knob to highlight the TRG_IN.WFM file.

￿Push ENTER to select the file.

3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.

4.Check external trigger high level:

a.Adjust oscilloscope controls: Press and hold the AWG2005 MANU+ AL TRIGGER button and adjust the oscilloscope vertical and hori: zontal position to display the waveform from the AWG2005. Release the MANUAL TRIGGER button.

b.Enable function generator output: Turn on the function generator output.

c.Check external trigger level accuracy:

￿Gradually increment the function generator offset level until a waveform is displayed on the oscilloscope.

￿Check that that the function generator offset level is from 0.85 to 1.159V when the waveform is first displayed.

5.Check external trigger low level:

a.Change the function generator controls:

Parameter

Offset-0.69V

b.Change the AWG2005 controls:

￿Push MODE￿Polarity to highlight Negative.

￿Select Level from the side menu, and turn the general purpose knob to select -19V. (You can also use the numeric and units keys to select -19V; then push ENTER.)

c.Check external trigger level accuracy:

￿Gradually decrease the function generator offset level until a waveform is displayed on the oscilloscope.

￿Check that that the function generator offset level is from -1.159V to -0.859V when the waveform is first displayed.

AWG2005 Service Manual

4￿41