Floating Point Processor Check (Option 09)

Performance Tests

This procedure checks the AWG2￿￿5 floating point processor￿

NOTE

This check requires that the AWG2005 has Option 09 installed.

Equipment Required: None￿

Prerequisites: The AWG2￿￿5 meets the prerequisites listed on page 4&7￿

Procedure:

1￿ Check that floating point processor test in internal diagnostics passes:

a￿ Run the AWG2005 internal diagnostics: Push the AWG2￿￿5 ON/ STBY switch two times so that the AWG2￿￿5 runs the power&on diagnostics￿

b￿ Check the FPP test results: When the AWG2￿￿5 finishes the FPP test￿ check that the test result is Pass￿

This completes the performance tests for the AWG2￿￿5￿

AWG2005 Service Manual

4￿49