Warranted Characteristics

This section will describe the warranted characteristics of the AWG2005. These can be divided into two main categories: electrical characteristics and environmental characteristics.

Performance

The electrical characteristics are valid under the following conditions:

Conditions

1) The instrument must have been calibrated at an ambient temperature

 

 

between +20_ C to +30_ C.

2)The instrument must be in an environment whose limits are described in Environmental Characteristics.

3)All tolerance limits apply after a 20 minute warm up and an execution of the self calibration.

4)The instrument is operating at an ambient temperature between

 

+10_ C to +40_

C, unless otherwise noted.

 

 

Electrical Characteristics

 

 

 

 

 

Characteristics

 

Description

Performance

 

 

 

Test

 

 

 

 

 

Clock Generator

 

 

Accuracy

 

 

+15_

C to +30_

C

0.005%

 

 

 

 

 

 

 

 

 

+10_

C to +40_

C

0.01%

 

 

 

Reference Oscillator

 

 

 

Accuracy

 

"1ppm

 

 

 

 

 

Stability

 

"1 ppm/year (20_ C to 30_ C)

 

 

 

 

 

 

 

 

 

Clock Generator (Option 05)

Check Clock Frequency Accuracy, page 4327.

 

Accuracy

 

 

 

 

 

 

 

 

+15_

C to +30_

C

0.0005%

 

 

 

 

 

 

 

 

 

 

 

+10_

C to +40_

C

0.001%

 

 

 

 

 

Reference Oscillator

 

 

 

 

 

 

 

 

 

Accuracy

 

"1ppm

 

 

 

 

 

 

 

Stability

 

"1 ppm/year (20_ C to 30_ C)

 

 

 

 

 

 

Check Clock Frequency Accuracy, page 4327.

AWG2005 Service Manual

1￿11