Performance Tests

Vertical

CH1

Coupling

DC

Scale

0.25V/div.

Input Impedance

505Ω

Horizontal

 

Sweep

5005μs/div.

Trigger

 

Source

CH1

Coupling

DC

Slope

Positive

Level

05mV

Mode

Auto

d.Set function generator controls:

Function

Square

Mode

Continuous

Parameter

 

Frequency

15MHz

Amplitude

2.05V

Offset

1.05V

Output

Off

2.Select the AWG2005 waveform file and set AWG2005 controls:

a.Initialize AWG2005 controls: Push UTILITY￿Misc￿Config...￿Re- set to Factory￿O.K.

b.Select waveform file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Turn the general purpose knob to highlight the EXT_CLK.WFM file.

￿Push ENTER to select the file.

3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.

4.Check the external CLOCK IN threshold level:

a.Enable function generator output: Turn on function generator output.

b.Check the level: Check that the waveform displayed on the oscillo6 scope has an amplitude of 5 divisions and a stable display of 5 cycles.

5.Turn off equipment output and disconnect test hookup:

a.Disable function generator output: Turn off function generator output.

b.Remove connections: Disconnect all connections to the AWG2005.

AWG2005 Service Manual

4￿43