Performance Tests
Vertical | CH1 |
Coupling | DC |
Scale | 0.25V/div. |
Input Impedance | 505Ω |
Horizontal |
|
Sweep | 5005μs/div. |
Trigger |
|
Source | CH1 |
Coupling | DC |
Slope | Positive |
Level | 05mV |
Mode | Auto |
d.Set function generator controls:
Function | Square |
Mode | Continuous |
Parameter |
|
Frequency | 15MHz |
Amplitude | 2.05V |
Offset | 1.05V |
Output | Off |
2.Select the AWG2005 waveform file and set AWG2005 controls:
a.Initialize AWG2005 controls: Push UTILITYMiscConfig...Re- set to FactoryO.K.
b.Select waveform file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the EXT_CLK.WFM file.
Push ENTER to select the file.
3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check the external CLOCK IN threshold level:
a.Enable function generator output: Turn on function generator output.
b.Check the level: Check that the waveform displayed on the oscillo6 scope has an amplitude of 5 divisions and a stable display of 5 cycles.
5.Turn off equipment output and disconnect test hookup:
a.Disable function generator output: Turn off function generator output.
b.Remove connections: Disconnect all connections to the AWG2005.
AWG2005 Service Manual | 443 |