Performance Tests
Press the bottom Configure button to select Slave.
4.Turn on CH1 output on both AWG2005 instruments : Push the CH1 button so that the LED above the CH1 output connector is on.
5.Check the slave operation:
a.Change the master AWG2005 operation mode from Triggered to Cont.
b.Check the operation: Check that the two waveforms displayed on the oscilloscope have a same timing relation and a stable display.
6.Disconnect test hookup:
a.Remove connections: Disconnect all connections to the both AWG2005 instruments and oscilloscope.
This procedure checks the AWG2005 DIGITAL DATA OUT at the rear panel.
NOTE
This check requires that the AWG2005 has Option 04 installed.
Electrical Characteristic Checked: Auxiliary Output, DIGITAL DATAOUT, Level, on page61714.
Equipment Required: Digital data out cable, 2 X13 header, probe and oscilloscope.
Prerequisites: The AWG2005 meets the prerequisites listed on page 477.
Procedure:
1.Install test hookup and set test equipment controls:
a.Hook up termination board: Connect a digital data output cable to the AWG2005 rear7panel CH1 DIGITAL DATA OUT output (see Figure 4721).
b.Hook up oscilloscope:
Connect the oscilloscope probe to the CH1 vertical input.
Connect the probe ground7clip to the GND pin of 2 X13 header.
446 | Performance Verification |