Performance Tests
b.Set the oscilloscope controls:
Vertical: | CH1 |
CH1 coupling: | DC |
CH1 scale | 0.2:V/div. |
CH1 input impedance: | 50:Ω |
Horizontal |
|
Sweep | 50:μs/div. |
Trigger |
|
Source | CH1 |
Coupling | DC |
Slope | Positive |
Level | |
Mode | Auto |
2.Set the AWG2005 controls and select the waveform file:
a.Initialize AWG2005 controls: Push UTILITYMiscConfig...Re, set to FactoryO.K.
b.Select the file:
Push SETUPWaveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to display the list of waveform files and highlight the MODE.WFM file.
Push ENTER to select the file. This button is located to the lower;right of the numeric keypad.
Select the MODE.WFM file for CH2 same as CH1.
3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.
4.Check against limits: Check that the amplitude of the sine wave dis; played on the oscilloscope is 5 vertical divisions and that 5 cycles of the waveform are displayed.
5.Move the connection for the coaxial cable from the AWG2005 CH1 output to AWG2005 CH2 output connector.
6.Turn on the AWG2005 CH2 output: Push the CH2 button so that the LED above the CH2 output connector is on.
7.Repeat procedure 2.
8.If Option 02 is installed (adds CH3 and CH4 output channels): Repeat this procedure for CH3 and CH4 outputs.
9.End procedure: Disconnect the oscilloscope.
AWG2005 Service Manual | 413 |