Performance Tests

b.Set the oscilloscope controls:

Vertical:

CH1

CH1 coupling:

DC

CH1 scale

0.2:V/div.

CH1 input impedance:

50:Ω

Horizontal

 

Sweep

50:μs/div.

Trigger

 

Source

CH1

Coupling

DC

Slope

Positive

Level

-100:mV

Mode

Auto

2.Set the AWG2005 controls and select the waveform file:

a.Initialize AWG2005 controls: Push UTILITY￿Misc￿Config...￿Re, set to Factory￿O.K.

b.Select the file:

￿Push SETUP￿Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).

￿Turn the general purpose knob to display the list of waveform files and highlight the MODE.WFM file.

￿Push ENTER to select the file. This button is located to the lower;right of the numeric keypad.

￿Select the MODE.WFM file for CH2 same as CH1.

3.Turn on the AWG2005 CH1 output: Push the CH1 button so that the LED above the CH1 output connector is on.

4.Check against limits: Check that the amplitude of the sine wave dis; played on the oscilloscope is 5 vertical divisions and that 5 cycles of the waveform are displayed.

5.Move the connection for the coaxial cable from the AWG2005 CH1 output to AWG2005 CH2 output connector.

6.Turn on the AWG2005 CH2 output: Push the CH2 button so that the LED above the CH2 output connector is on.

7.Repeat procedure 2.

8.If Option 02 is installed (adds CH3 and CH4 output channels): Repeat this procedure for CH3 and CH4 outputs.

9.End procedure: Disconnect the oscilloscope.

AWG2005 Service Manual

4￿13