Calibration

Self Tests

The AWG2005 includes internal calibration routines that check electrical characteristics such as amplitude, offset, trigger level, clock, filters and attenuation and adjust internal calibration constants as necessary. This procedure describes how to do the internal calibration.

Equipment Required: None.

Prerequisites: Power on the AWG2005 and allow a 20 minute warmup period at an ambient temperature between +15°C and +25°C before doing this procedure.

Procedure:

NOTE

￿f the AWG2005 is powered off while the calibration is executed the calibration data in the memory may be loss.

1.Verify that internal adjustments pass: Do the following substeps to verify internal adjustments have passed.

a.Display the calibrations menu and select all tests: Push UTILITY￿Diag/Cal￿Calibrations xxxx￿All. See the menu in Figure￿402.

The Calibrations column on the left shows the tests available for calibration. In addition to selecting all of the tests shown, you can select only the test(s) you want to run using the general purpose knob. In Figure 402, the symbol to the left of Trigger indicates the tests selected.

AWG2005 Service Manual

4￿5