Performance Tests

Pulse Response Check

7. End procedure: Disconnect the DMM.

This procedure checks the pulse response characteristics of the AWG2005 output waveforms at amplitudes of 0.5 and 15V.

Electrical Characteristic Checked: Main Output, Pulse Response, on

page51612.

Equipment Required: A 505Ω coaxial cable and an oscilloscope.

Prerequisites: The AWG2005 meets the prerequisites listed on page 467.

Procedure:

1.￿nstall test hookup and set test equipment controls:

a. Hook up the oscilloscope: Connect the AWG2005 CH1 output connector through the coaxial cable to the CH1 vertical input con6 nector on the oscilloscope (see Figure 4615).

AWG2005

Oscilloscope

 

 

 

AWG2005

Figure 4#1￿:"Pulse Response Initial Test Hookup

b.Set oscilloscope controls:

Vertical

CH1

Coupling

DC

Scale

0.15V/div.

Input impedance

505Ω

Horizontal

 

Sweep

205ns/div.

Trigger

 

Source

CH1

Coupling

DC

Slope

Positive

Level

05V

Mode

Auto

4￿34