Performance Tests
Flatness | 0.157div., maximum |
Fall time | 357ns, maximum |
6.Move the connection for the coaxial cable from the AWG2005 CH1 output to AWG2005 CH2 output connector.
7.Turn on the AWG2005 CH2 output: Push the CH button so that the LED above the CH2 output connector is on.
8.Repeat procedures 4 through 5.
9.If Option 02 is installed (add CH3 and CH4 output channels: Repeat above procedure for the AWG2005 CH3 and CH4 output channels.
10.End procedure: Remove the connections.
MARKER OUT | This procedure checks the amplitude of the MARKER OUT signal. |
Amplitude Check |
|
| Electrical Characteristic Checked: Auxiliary Output, MARKER, Ampli8 |
| tude, on page71814. |
| Equipment Required: A 507Ω coaxial cable and an oscilloscope. |
| Prerequisites: The AWG2005 meets the prerequisites listed on page 487. |
Procedure:
1.Install test hookup and set test equipment controls:
a.Hook up the oscilloscope: Connect the AWG2005 rear8panel CH1 MARKER OUT connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope (see Figure 4816).
AWG2005 | Oscilloscope |
AWG2005
Figure 4)16:(Initial Test Hookup
b.Set oscilloscope controls:
436 | Performance Verification |