Performance Tests
AWG 2005 Rear Panel | Oscilloscope |
|
213 Header
Digital Data
Out Cable
Figure 4221:1Digital Data Out Initial Test Hookup
c.Set oscilloscope controls:
Vertical | CH1 |
Scale | 11V/div. |
Input Impedance | 11MΩ |
Horizontal |
|
Sweep | Adjust as needed |
Trigger |
|
Mode | Auto |
2.Create the AWG2005 waveform file, select waveform file, and set AWG2005 controls:
a.Initialize AWG2005 controls: Push UTILITY!Misc!Config...!Re2 set to Factory!O.K.
b.Select waveform file:
Push SETUP!Waveform Sequence, if necessary, to select a waveform file for CH1. Waveform Sequence toggles between the CH1 files (upper list) and the CH2 files (lower list).
Turn the general purpose knob to highlight the DIGI_OUT.WFM file.
Push ENTER to select the file.
c.Set AWG2005 controls:
Push MODE!Cont
3.Check the CH1 digital data output signals:
AWG2005 Service Manual | 4!47 |