erformance Tests
This subsection contains a series of procedures for checking that the AWG2005 Arbitrary Waveform Generator performs as warranted.
The procedures are arranged in eleven logical groupings, presented in the following order:
Operating Mode Checks
Arithmetic Operation Checks
Clock Frequency and Amplitude Checks
Gain Accuracy Check
Offset Accuracy Check
Pulse Response Check
MARKER OUT Amplitude Check
CONTROL SIG OUT Amplitude Check
External Trigger Level Accuracy Check
External CLOCK IN Check
DIGITAL DATA OUT Check
These procedures extend the confidence level provided by the internal diagnostic and calibration routines described on page 4=3.
The tests in this subsection comprise an extensive, valid confirmation of performance and functionality, when the following requirements are met:
You must have performed and passed the calibration procedure de= scribed in Self Tests, the previous subsection.
NOTE
For operation to specified accuracy, allow the AWG2005 to warm up at least 20 minutes before doing the performance tests.
Load all the files from the Performance Check disk (063=1706=XX) that comes with this manual into AWG2005 internal memory. For instructions on loading files, see Loading Files on page<2=12 in the Instructions for Operation subsection in section 2.