Performance Tests

Function Generator

FG 5010

AWG2005

AWG￿005

DMM

884￿A

Figure 419:0External AM Operation Initial Test Hookup

c.￿et DMM controls:

Mode

VDC

Range

￿0

Inputs

Front

d.￿et function generator controls:

Function

Square

Mode

Continuous

Parameter

 

Frequency

1/kHz

Amplitude

0/V

Offset

1/V

Output

Off

￿. ￿et the AWG2005 controls and select the waveform file:

a.Initialize AWG2005 controls: Push UTILITY￿Misc￿Config...￿Re1 set to Factory￿O.K.

b.Modify AWG2005 default settings:

￿Push MODE￿Autostep.

c.￿elect waveform file:

￿Select Select Autostep File from the side menu.

￿Turn the general purpose knob to highlight the OPE.AST file.

￿Push ENTER to select the file.

AWG2005 Service Manual

4￿23