Performance Tests

CONTROL SIG OUT Amplitude Check

This procedure checks the amplitude of the CONTROL SIG OUT signal.

Electrical Characteristic Checked: Auxiliary Output, CONTROL SIG,

Amplitude, on page 19148.

Equipment Required: A 508Ω coaxial cable and an oscilloscope.

Prerequisites: The AWG2005 meets the prerequisites listed on page 497.

Procedure:

1.Install test hookup and set test equipment controls:

a.Hook up the oscilloscope: Connect the AWG2005 rear9panel CON9 TROL SIG OUT connector through the coaxial cable to the CH1 vertical input connector on the oscilloscope (see Figure 4917).

AWG2005

Oscilloscope

AWG2005

Figure 4/17:.Initial Test Hookup

b. Set oscilloscope controls:

 

Vertical

CH1

CH1 Coupling

DC

CH1 Scale

18V/div.

CH1 Input Impedance

508Ω

Horizontal

 

Sweep

508ns/div.

Trigger

 

Source

CH1

Coupling

DC

Slope

Positive

Level

5008mV

Mode

Auto

2.Set the AWG2005 controls and select the waveform file:

a.Initialize AWG2005 controls: Push UTILITY￿Misc￿Config...￿Re/ set to Factory￿O.K.

4￿38

Performance Verification