Performance Tests

3.End procedure: Remove equipment: Disconnect connections to the test equipment.

Clock Frequency and These procedures check the accuracy of the AWG2005 clock frequency and

Amplitude Checks

the waveform output amplitude.

 

 

Check Clock Frequency Accuracy

 

Electrical Characteristic Checked: Clock Generator, Accuracy, on

 

page01111.

 

Equipment Required: A 500Ω coaxial cable and a frequency counter.

 

Prerequisites: The AWG2005 meets the prerequisites listed on page 417.

Procedure:

1.￿nstall test hookup and set test equipment controls:

a.Hook up frequency counter: Connect the AWG2005 rear panel CLOCK OUT connector to the frequency counter input through a coaxial cable (see Figure 4111).

Frequency Counter

DC 5010

AWG2005

AWG2005

Figure 4+11:*Clock Frequency Accuracy Initial Test Hookup

AWG2￿￿5 Service Manual

4￿27