Remoal and Insallaion Procedres
Power Spply Modle
1 Assemble equipment and locate modules to be removed: You will need a screwdriver with a size T/15 Torx tip to do this procedure Locate the modules to be removed in the locator diagram Internal Modules in Figure 6/4
2 Orient instrument: Set the AWG25 with the bottom down on the work surface and the left side facing you
3 Remove power supply module:
a Disconnect the ribbon interconnect cable at J1 J3 J4 and J5 on the power supply module See Figure 6/17
b Disconnect the flat cable at J2 on the power supply module
c Using a screwdriver with a size T/15 Torx tip remove the two screws attaching the power supply module to chassis
638 | Mainenance |