Removal and Installation Procedures

Connector Module

1.Assemble equipment and locate modules to be removed:

a.You will need a screwdriver with a size T79 Torx tip and a 1@27inch nut driver to do this procedure.

b.Locate the modules to be removed in the locator diagram￿ Internal Modules￿ in Figure 674.

2.Remove front#panel knob: Do the Front#panel Knob procedure￿ on page66717.

3.Remove front cover, trim ring, and menu buttons: Do the Front Cover, Trim Ring, and Menu Buttons procedure on page 6722.

4.Orient instrument: Set the AWG2005 with the top down on the work surface and the right side facing you.

5.Remove connector module:

a.Disconnect the interconnect cables at the CH1and CH2 Waveform Output connectors. For an AWG2005 with Option 02￿ disconnect the interconnect cables at the CH3 and CH4 Waveform Output connec7 tors.

b.Using the screwdriver with a size T79 Torx tip￿ remove the four screws attaching the connector module to the chassis. See Figure 6713.

6.Remove BNC connector:

a.To remove a BNC connector￿ remove the nut attaching the BNC connector to the panel using a 1@27inch nut driver￿ and then remove the BNC connector from the panel.

7.Reinstallation:

a.Install BNC connector: Do substep 6a￿ reversing the order of the items in the substep.

b.Install connector module: Install the connector module by doing substeps 5a-5b in reverse order.

￿WG￿￿￿￿ Service Manual

6￿31