Verifying Performance

Agilent Technologies 8922M/S Specifications

AC Voltage Measurement

Voltage Range

Accuracy (20 Hz to 15 kHz), Input >1 mVrms

Residual Noise + THD (15 kHz BW) Supplemental Characteristics

3 dB Bandwidth

Input Impedance

Minimum Resolution

0 V to 30 Vrms

±3% of reading

175μV

2 Hz to 100 kHz

1 MΩ 145 pF at AUDIO IN

4 digits for inputs 100mV

3 digits for inputs <100 mV

DC Voltage Measurement

Voltage Range

Accuracy

DC Offset:

Supplemental Characteristics Minimum Resolution

100 mV to 42 V

±(1.0% of reading + DC Offset)

±45 mV

1.0 mV

Distortion Measurement

Fundamental Frequency

Input Level Range

Display Range

Accuracy

Residual THD + noise (15 kHz BW) Supplemental Characteristics

Minimum Resolution

1 kHz ±5 Hz

30 mVrms to 30 Vrms

0.1% to 100%

±1 dB (0.5 to 100% distortion) The greater of 60 dB or +175 μV

0.01% distortion

3-26

Page 101
Image 101
Agilent Technologies 8922M, S GSM manual AC Voltage Measurement, DC Voltage Measurement, Distortion Measurement