Connectors

Front-Panel Connectors of the Agilent Technologies 8922M/S

Front-Panel Connectors of the Agilent Technologies 8922M/S

1. AUX RF IN

The auxiliary RF input connects to the input section and to the RF analyzer (if

 

selected). This connector provides a higher sensitivity and lower maximum-power

 

connection from the DUT. It is not normally used for transceiver testing but is a

 

useful input to the Spectrum Analyzer (option 006).

 

AUX RF IN has its own input level setting (RF analyzer amplitude) separate from

 

RF IN/OUT. It is reverse power protected.

 

This connector is selected when the RF analyzer’s RF input is set to AUX RF IN.

 

See Also

Screens: RF Generator/RF Analyzer, (RF Analyzer)

2.AUX RF OUT The auxiliary RF output connects to the output section from the RF generator’s step attenuators (if selected). This connector provides higher output levels to a device- under-test (DUT). It is not normally used for transceiver testing.

This connector has its own output amplitude setting (RF generator amplitude) separate from RF IN/OUT. It is reverse power protected.

This connector is selected when the RF generator’s RF output is set to AUX RF

OUT.

See Also

Screens: RF Generator/RF Analyzer, (RF Gen)