Messages

Sync Status

 

Sync Status

 

This field displays any errors that occurred while trying to synchronize the

 

demodulated data. Some screens use this field to display progress during multi-burst

 

measurement. When the measurement is completed to field returns to its normal

 

state.

Bad Sync (for

This message appears if a synchronization error happened since the last

demodulation only)

Demodulation Arm (when Adjust Mode is disabled) or since the last change to Trig

 

Delay (when Adjust Mode is enabled with Demod Arm already selected).

 

Possible causes of Bad Sync are:

 

• Useful bits occurred while power was too low.

 

• Demodulation trigger too early or too late.

 

• FM (bit) errors found while synchronizing to desired midamble.

 

• RF overload.

FM Error

At least one bit error was detected when comparing the measured midamble to the

 

selected Midamble or User Defined Sync Pattern, (Sync Mode=Midamble only).

Level Late

The amplitude of the burst did not rise until after the first few bits were received.

Level Short

The amplitude of the burst fell before the last few bits were received.

Low Level

The DSP analyzer’s RF level did not rise high enough to make a valid measurement.

No Error

No synchronization error occurred.

RF Ovrload

The measurement hardware overloaded during the measurement. (Increase RF

 

Analyzer Amplitude setting to correct).

ShortBurst

The amplitude envelope was shorter than the expected burst.