Screens

 

 

Phase and Frequency Error - Multiburst OFF

3. ON/OFF

This field enables or disables multi-burst measurement. The default for this field is

 

OFF.

 

 

Choices

ON enables multi-burst measurement. The display changes to

 

 

show Bursts measured fields; Wanted, Done, and Errors.

 

 

OFF disables multi-burst measurement.

 

See Also

Screens: Phase and Frequency Error - Multi-burst ON

4.Peak Phase Error This field displays the measured Peak Phase Error over the useful bits in the measured burst.

See Also

Keys: ON/OFF, HI LIMIT, LO LIMIT, REF SET, AVG,

 

METER, Units

5.RMS Phase Error This field displays the measured RMS Phase Error over the useful bits in the measured burst.

 

 

See Also

Keys: ON/OFF, HI LIMIT, LO LIMIT, REF SET, AVG,

 

 

 

METER

6.

Single/Cont

Single allows one test to be performed. Cont allows testing to automatically repeat.

7.

Sync Status

This field displays any errors that occurred while trying to synchronize to the

 

 

demodulated data.

 

 

See Also

Making Measurements: Solving Problems Messages

8.

Trig Delay

This field sets the time delay between a valid trigger event and the beginning of a

 

 

measurement.

 

 

 

Range

0 to 5000.00 μs.

9.

View

This field selects alternate views of the measurement.

 

 

Choices

Phase Err

 

 

 

Data Bits

4-87

Page 193
Image 193
Agilent Technologies 8922M, S GSM manual On/Off