Agilent Technologies S GSM, 8922M manual Loading, Storing, and Running

Models: 8922M S GSM

1 381
Download 381 pages 1024 b
Page 306
Image 306

Instrument BASIC

Loading, Storing, and Running

Loading, Storing, and Running

This section describes loading, storing, and running both IBASIC programs and test procedures using the TESTS subsystem.

Loading An IBASIC

1

Insert the memory card.

Program From A

2

Access the IBASIC Controller screen from the Test Function field on the

Memory Card

 

TESTS screen.

 

 

 

3

Using the knob, select the field and enter the following command to load your

 

 

program:

 

 

GET “<filename>:INTERNAL”

Loading an IBASIC

1

Insert the disk into the drive.

Program From A

2

Access the IBASIC controller screen from the Test Function field on the TESTS

GPIB Disk Drive

 

screen.

 

 

 

3

Using the knob, select the field and enter the following command to load your

 

 

program:

Downloading An

IBASIC Program

Into the

Agilent 8922M/S

LOAD “<filename>:7xx,x”

This procedure downloads an IBASIC program from your connected IBASIC computer to the Agilent 8922M/S’s IBASIC Controller. This procedure assumes your Agilent 8922M/S’s GPIB address is set to 14.

1Access the Agilent 8922M/S’s IBASIC Controller screen.

2Load your IBASIC program into your connected IBASIC computer.

3Enter the following commands on your IBASIC computer to copy the program into the Agilent 8922M/S’s IBASIC Controller:

OUTPUT 714;”PROG:DEL” <enter>

OUTPUT 714;”PROG:DEF #0” <enter>

LIST #714 <enter>

OUTPUT 714;” “END <enter>

8-10

Page 306
Image 306