Agilent Technologies 8922M, S GSM manual Cell Control, Single Other Settings

Models: 8922M S GSM

1 381
Download 381 pages 1024 b
Page 143
Image 143

Screens

Cell Control 2

Cell Control 2

68

5

4

1

3

 

 

 

7

2

 

 

 

 

1.

Adj Cell

This field selects which adjacent cell SACCH measurement result to display. The

 

 

adjacent cells are prioritized by power level. One is the largest signal.

 

 

Range

 

1 to 6

2.

Amplitude

This field changes the RF generator amplitude. It is a duplicate of the RF generator

 

 

Amplitude field on the RF Generator/RF Analyzer screen.

 

 

See Also

 

Screens: RF Generator/RF Analyzer (RF Generator).

3. ARFCN

This field selects the ARFCN for TCH1 or TCH2 when the Mode field is set to

 

 

Single.

 

 

 

 

Other Settings

 

MA1 or MA2 appears in this field when Mode is set to Hopped.

4.

Bad Syncs

This field is a count of bad synchronization occurrences since the last reset.

4-37

Page 143
Image 143
Agilent Technologies 8922M, S GSM manual Cell Control, Single Other Settings