Hardware

Connections and

Agilent 8922M/S

Configuration

Instrument BASIC

Configuration and Instrument Control

Agilent 8922M/S provides an RS-232 Serial port and an GPIB port for a variety of uses:

Controlling the Agilent 8922M/S using a connected controller

Controlling connected instruments using the Agilent 8922M/S IBASIC computer

Printing screen images and test results

Entering and editing IBASIC programs

The Agilent 8922M/S’s I/O CONFIGURE screen is used to configure these ports for the desired use.

GPIB Configuration 1 Access the Agilent 8922M/S’s I/O CONFIGURE screen.

For Programming

2Set the HP-IB Adrs to the desired address.

3Set the Mode field to Talk&Lstn.

4Enter the Print Adrs if a GPIB printer is connected. 1 Set the Printer Port field to HPIB.

5Connect GPIB cables to other instrument(s).

This configuration prepares the Agilent 8922M/S to be controlled by a system controller, allowing program transfers over the bus. Refer to ‘Entering and Editing Programs” on page 8-16, to start programming or editing.