Cell Control - CW Generator

Cell Control - CW Generator

 

This screens allows you to set the Agilent 8922M/S up as a CW Signal Generator.

1. Base Station

There are two Base Station fields. The Amplitude field sets the amplitude of the

 

Agilent 8922M/S transmission. The Channel field selects which channel the

 

Agilent 8922M/S transmits the Broadcast Control Channel (BCCH) information. If

 

this is changed during a call, the call will be deactivated. The frequency can also be

 

set explicitly between 10 MHz and 1000 MHz.

2.Burst Type. This field defines the expected burst type of the mobile phone’s transmitted bursts. When the Traffic Channel (TCH) is selected, the correct Colour Code should be chosen.

3. Channel

This field displays the channel number the Agilent 8922M/S is taking measurements

 

from. For a GSM900 mobile phone the channel number may vary from 1 through

 

124. For an E-GSM mobile phone the channel number may vary from 0 through 124

 

and 975 through 1023. If you change the channel number, the frequency field value

 

also changes. The converse of this is not the case. You may enter a non GSM

 

standard channel frequency on another screen, between 10 MHz and 1000 MHz.