Screens

 

 

Cell Control - Active Cell

13. Pwr Ramp

This field displays the power mask and, (option 006 only), pulse measurements

 

 

screens.

14. Pwr Zero

This field zeros the power meter. RF power must be disconnected from the RF

 

 

IN/OUT port when executing this function.

15.

Reset

This field resets all the Agilent 8922M/S’s measurements.

16.

Scope

This field displays the oscilloscope measurements screen.

17.Spec Anl (Option This field displays the spectrum analyzer measurements screen.

006 only)

18.Mobile Reports The mobile reports reflect the status of the signal being received by the mobile phones receiver.

19.TX Level/Manual This field sets the expected input level of the RF Analyzer on the RF IN/OUT port. This level must be within ±3 dB of the actual level in order that fully calibrated

measurements may be made.

This field displays the amplitude the Agilent 8922M/S is expecting from the mobile phone, and adjusts its RF analyzer setting to match this input. When this field is changed, the mobile phone TX Level and the Agilent 8922M/S RF Analyzer are no longer coupled together. This means that you have independent control of the amplitude of the transmitted mobile phone signal and the sensitivity of the Agilent 8922M/S analyzer circuit.

When the fields are coupled, the Expected Input shows the power level of the setting. When they are independent, the Expected Level displays Manual.

Range

GSM900 and

27.9 to +41.0 dBm.

 

E-GSM:

 

The fields in the Mobile Phone area display information about

 

the call.

 

See Also

Chapter 3: Specifications

20.Channel (Mobile) The channel field allows you to change the channel number of the call, either before or during the call. If this field is changed when a call is in progress, a handover is performed.

4-29

Page 135
Image 135