Verifying Performance

Installing and Operating the Software

Forward Conversion

To return the instrument from an Agilent 8922G back to an Agilent 8922M or an

Agilent 8922E to an Agilent 8922S, select the following keys:

More (this is accessible from the Cell Control screen in the bottom right-hand corner). Scroll down the list and select CONFIG.

Compatible, select 8922M or 8922S

HP-IB Adrs (14)

PRESET

The instrument is returned to an Agilent 8922M or Agilent 8922S.

To Configure the

1 With the program loaded, type EDIT DEFAULT_ADDRESS, press

ENTER

.

 

GPIB Addresses

 

 

 

2Modify each line to indicate the proper instrument address (700-730).

It is now possible to re-store the program as “PT_8922” or store it under a different name.

To Run the Program 1 Type RUN, press ENTER .

2Follow the directions as they appear on the screen.

Notes on Running the Program.

The first screen which appears is the GPIB status of each piece of test equipment that is supported. It is only necessary to have the instruments responding that will be used in each particular test. Make certain that each instrument you will be using is responding at the proper address. Duplicate addresses may make an instrument appear to be responding but this is not allowed. Press “I” (for Ignore) to continue past this screen.

The second screen prompts you for the instrument model. If you have disk

08922-10006, select HP 8922G (for Agilent 8922M performance testing) or HP 8922E (for Agilent 8922S performance testing). The third screen which will appear is the main Performance Tests selection menu. Three options are available on this screen:

Select the performance test to run, remember the test instruments and UUT must be responding over GPIB.

Turn the printer function ON or OFF. If the printer function is turned on it must be responding over GPIB or the program will lock up.

Exit from the program.

Press the key corresponding to the option that you would like to perform. The other screens that appear are connection instructions, error messages and output results.

3-6

Page 81
Image 81
Agilent Technologies 8922M, S GSM manual Forward Conversion, Gpib Addresses, To Run the Program 1 Type RUN, press Enter