Pulse Measurements (Available if Option 006, Spectrum Analyzer, is Fitted)

Making Measurements

Measurements

If you have option 006 (spectrum analyzer) installed, you can make measurements on the lower portion of the pulse. These measurements can be accessed from the Power Ramp screens.

The main measurements, Pulse On/Off Rise and Pulse On/Off Fall are displayed at the top of the screen (1), see Figure 2-9 on page 2-18

The sensitivity of the Agilent 8922M/S receiver can be varied (2) allowing you to verify whether the signal still falls within the boundaries of the GSM power mask.

1

3

2

Figure 2-9

Pulse Measurements

The Agilent 8922M/S provides three pulse measurements which can be selected using View (3). These are:

Pulse - pulse on/off ratio measures the ratio of the average transmitter- power

(pulse on) to a specified time-position when power is reduced (pulse off). The default settings are 28μs before bit 0 and 28μS after bit 147.

Puls Rise - this screen displays the signal during the initial rise of the pulse.

Puls Fall - this screen displays the signal during the final fall of the pulse.

2-18