Screens

 

RF Generator / RF Analyzer (RF Analyzer)

10. Hop Trig

This field selects whether the RF Analyzer is armed or disarmed to accept a hop

 

trigger. Hop Trig cannot be set to Arm until Hop Mode is set to Hop.

 

 

NOTE

Do not make measurements with Hop Trig set to Disarm and Hop Mode set to Hop.

 

 

11.Open/Auto DAC This field is the AGC level DAC value when in Open or Auto AGC mode. This value

Value

can be entered automatically by choosing Auto AGC Mode. It can also be entered

 

by performing a Do Open Cal operation while in Open AGC mode, or manually

 

while in Open or Closed AGC Mode.

 

During Auto operation, the DAC value is based on the RF Analyzer Amplitude

 

setting, and will be updated each time RF Analyzer Amplitude is changed.

 

See Also

Screens: RF Analyzer (Do Open Cal, AGC Mode)

 

Range

0 to 255.

12. RF Input

This field selects the RF input port for the RF Analyzer.

 

Choices

RF IN/OUT

 

 

AUX RF IN

 

Range

RF IN/OUT: 27.9 to +41.0 dBm. AUX RF IN: 58.0 to +20.0

dBm.

4-114