Agilent Technologies S GSM, 8922M Test Execution Conditions, To Continue or Pause After Each Test

Models: 8922M S GSM

1 381
Download 381 pages 1024 b
Page 326
Image 326

Instrument BASIC

Programming and Using the TESTS Subsystem

1

2

3

5

4

Figure 8-12

Test Execution Conditions

To Have Testing Stop or Continue on a UUT Failure

Refer to item (1) in Figure 8-12 on page 8-30.

On UUT Failure.….….…. Continue

Testing continues whenever the UUT (Unit Under Test) fails to meet its test specification limits. When this occurs, an error is listed on the test-results printout and/or is displayed on the CRT.

On UUT Failure.….……Stop

Testing stops whenever the UUT fails to meet test specification limits.

To Continue or Pause After Each Test

Refer to item (2) in Figure 8-12 on page 8-30.

Run Mode.….……Continuous

All tests run in sequence. Testing pauses only if the operator is required to interact with the UUT or Agilent 8922M/S; interaction such as changing UUT channels, setting squelch, changing audio level, and so forth, cause testing to pause.

Run Mode.….……Single Step

The program stops running at the completion of each test. The test-system operator is prompted to select Continue to proceed with testing.