Configure

8. I/O Config

This field allows you to access the I/O configuration screen. This screen allows you

 

to select your printer configuration, GPIB and serial port communication settings.

9.Meas Reset (Meas Selecting Meas Reset will erase any accumulated measurements used for calculating

Cntl)

a final result, and re-start the measurement process for the following functions:

 

HI LIMIT

 

LO LIMIT

 

AVG

10. Offset

This field sets the reference frequency offset in parts-per-million (ppm).

11. OPT 001

If installed, this field turns the optional high stability timebase on or off.

REF OUT

 

 

12.Radio Type This field selects the type of radio to be tested.

Choices

GSM 900

 

E-GSM

 

DCS 1800

 

PCS 1900

13.Range Hold This field selects the ranging mode for the AF analyzer and the spectrum analyzer.

Choices

Auto All selects auto ranging.

 

Hold All selects range hold, allowing manual setting of spectrum

 

analyzer input attenuation and AF Analyzer gain control.

See Also

Screens: Spectrum Analyzer (Input Atten)

4-51