Verifying Performance

Agilent Technologies 8922M/S Specifications

Spectral Purity

Spurious Signals (for +1 dBm output level at Aux RF Out or ≤−19 dBm output level at RF In/Out.

Harmonics:

<25 dBc

Non-harmonics:

<50 dBc, >5 kHz offset from carrier

 

 

0.3 GMSK Modulation

After one timeslot, 577μs, from an isolated RF Generator Trigger in the GSM Frequency bands.

Phase Error

1° rms

Peak Phase Error

4° peak

Frequency Error

± [0.02 ppm (18 Hz) + reference accuracy], for normal bursts

 

Typically ±[0.03 ppm (27 Hz) + reference accuracy], for RACHs.

Amplitude Flatness

±0.25 dB peak.

Clock Input (Agilent 8922M only)

 

Frequency

270.833 kHz ± 2 Hz (relative to reference).

Level

TTL

Data Input (Agilent 8922M only)

 

Format

Non differentially encoded input.

Level

TTL

 

 

Supplemental Characteristics:

After three timeslots, 1.73 ms, from an isolated RF Generator Trigger in the GSM frequency bands

Phase Error

0.5° rms

Peak Phase Error

2.0° peak

Frequency Error

±[0.01ppm (9 Hz) + reference accuracy] for normal bursts.

 

±[0.02ppm (18 Hz) + reference accuracy] for RACH bursts.

 

 

Pulse Modulation

Input Levels (Agilent 8922M only)

Rise/Fall Time (10% to 90%)

Supplemental Characteristics:

On/Off Ratio

TTL 5 μs

>80 dB

3-16

Page 91
Image 91