Connectors

Signal Descriptions for SYSTEM BUS

RP_TXD, RP_RXD

TXD-Pin 37 RXD-Pin 18 Output/Input

 

These are the rear-panel serial transmit data and rear-panel serial receive data lines.

 

 

NOTE

Ground signals not listed individually here.

 

 

 

RP_TX_HOP

Pin 28 Input

 

 

This is the rear-panel trigger signal input for hopping the RF generator (if selected).

 

It is used when externally addressing the hop frequencies. It is connected to the hop

 

controller. It can also be configured to control RF generator output pulsing

 

(automatic level pulsing when this trigger occurs).

 

You must supply signals on the rear-panel hop frequency table address input lines to

 

select each RF generator hop frequency for each hop trigger. Also, the hop

 

controller’s hop address Source field must be set to Ext.

 

This line is active when the hop controller’s Source field is set to Ext, the RF

 

generator’s hop mode is set to Hop, and the RF generator’s hop trigger is set to Arm.

 

Requirements

Amplitude: TTL levels High drive requirement: 100 μA Low

 

 

drive requirement: 2 mA Triggered by: rising edge

 

See Also

Screens: RFG/RFA (Hop Control) Specifications Timing

 

 

Diagrams

SEQ_TRIG_OUT

Pin 9 Output

 

 

This is the sequence trigger output. It is connected to the DSP analyzer. It outputs a

 

positive-going pulse when an internal sequence goes through something other than

 

the next higher address, or when reset occurs (Hop Address Source=Int).