Making Measurements

Advanced Features

Unexpected

These are descriptions of operations which may be unclear to the user as to how they

Operations

occurred. These are NOT defects.

 

1

The Agilent 8922M/S occasionally sends ABORTS to the attached protocol mon-

 

 

itor. These ABORTS are normal and logging data is not lost.

 

2

In the FA and SD/4+FA control channel configurations, the Agilent 8922M/S

 

 

sends speech frames rather than fill frames on the FACCH when there is nothing

 

 

else to send.

 

3

The DRX ON mode corresponds to the NORMAL paging mode. The

 

 

Agilent 8922M/S sends continuous paging mode NORMAL on the idle PCH

 

 

blocks. The DRX OFF mode corresponds to the REORGANIZE paging mode.

 

 

The Agilent 8922M/S sends continuous paging mode REORGANIZE on the idle

 

 

PCH blocks. When in the REORGANIZE paging mode, the Agilent 8922M/S is

 

 

guaranteed to page on a random PCH block which is not the block expected in

 

 

the NORMAL paging mode. The active PCH block will page the MS with the

 

 

paging mode set to SAME. The SAME mode means “the same paging mode as

 

 

last indicated.”

 

4

TMSI reallocation in the Agilent 8922M/S uses the two (2) octet TMSI size.

 

5

The Agilent 8922M/S does not turn off the transmitter during DTX when the

 

 

voice activity detector (VAD) determines that there is no speech input present.

 

 

Instead, the Agilent 8922M/S sends a silence descriptor (SID) on every speech

 

 

frame.

 

6

USE MEM is not intended to be used with Pulse (Pulse ON/OFF Ratio)

 

 

measurements - do not select USE MEM after making a pulse measurement.

 

7

During a Pulse (Pulse ON/OFF Ratio) measurement the following message can

 

 

occur if the device under test is generating a larger signal than is expected: ‘RF

 

 

Overpower occurred during Pulse On/Off Ratio measurement’. This message

 

 

cannot be queried via GPIB. Use the Hardware 1 Status Register Bit 5 to detect

 

 

this condition when doing pulse measurements.

 

 

NOTE

Option 006 (spectrum analyzer) must be installed to make pulse measurements.

 

8

The Output RF Spectrum measurement and Pulse On/Off Rise and Fall

 

 

 

measurement results shown on the Main view screens are not accessible via

 

 

GPIB or IBASIC. These measurement results are available on the Trace view

 

 

screens. All GPIB and IBASIC actions with these fields refer to those fields

 

 

shown on the Trace view screens.