Making Measurements

 

Advanced Features

 

The beginning of the 0 sample bin will be between 0 and 75μS (bit 0 - 20) after the

 

trigger as the sampler runs asynchronously to the trigger. The last sample bin will

 

complete sampling between 450μS and 525μS (bit 121 - 142) into the burst.

 

The measurement processor discards the 0 sample bin. This effectively removes the

 

ramp-up overshoot. Each measurement is also RF Power Qualified to ensure only

 

valid bursts are measured. Measurements with a result below the qualification level

 

are discarded.

 

The remaining sample bins 1-5 are averaged for the displayed measurement result.

 

Each measurement is therefore based on a 375μS (102 bit) sampling of the burst.

 

A mobile which does not meet the amplitude mask and has erratic amplitude across

 

the burst could give erroneous results with this measurement algorithm. It is

 

therefore essential that mobiles first comply with the amplitude mask in

 

GSM Rec. 11.10.

Making

There are two ways of making a synchronized external measurement using triggers

Synchronized

from the Agilent 8922M/S:

External

1 Use the G_EXT_TRIG on the rear panel SYSTEM_BUS connector. See

Measurements Using

Triggers From The

Chapter 6, Connectors for a pin out diagram.

Agilent 8922M/S

This is a separate trigger directly from the Protocol Processor. It is only

 

 

selectable from the SERVICE screen. Once set it will provide continuous

 

triggers on each frame. Caution must be used as this trigger can have up to a

 

0.75 bit error due to the 156/157 bit variance in timeslots.

 

(GSM Recommendations allow use of 156/157 bit timeslots instead of the

 

156.25 bit length.)

 

To use the G_EXT_TRIG signal:

 

 

Go to the CONFIG screen.

 

 

Select the SERVICE field on the right hand side of the screen.

 

 

Select the LATCH field and find the selection labeled:

 

 

‘g_ext_trig_enable’ (SERV:LATCH:SEL ‘g_ext_trig_enable’)

 

 

Enter a ‘1’ in the VALUE field. (SERV:LATCH:VAL 1)

 

 

Select the LATCH field and find the selection labeled:

 

 

‘g_ext_trig’ (SERV:LATCH:SEL ‘g_ext_trig’)

 

 

Enter the TIMESLOT and BITPOSITION on which you want the

 

 

trigger to occur.

 

Example:

432 hex is TIMESLOT=4 BITPOSITION=32 hex. The

 

 

maximum TIMESLOT=7 and the maximum BITPOSI-

 

 

TION=9C hex (Bit 0 - 157).

2-42