Agilent Technologies 8922M, S GSM manual 11/19. Ratio/Count, Auto

Models: 8922M S GSM

1 381
Download 381 pages 1024 b
Page 119
Image 119

 

 

Screens

 

 

Bit Error 2

11/19. Ratio/Count

These fields select how frame errors/CRC errors will be displayed for both the

 

Intermediate and Completed results.

 

Choices

Error displays the frame errors/CRC as a percentage per

 

 

measurement.

 

 

Count calculates and displays the total number of bit errors

 

 

counted to the total number of bits measured (depends on

 

 

measurement type).

 

Field (11) displays the final result of the measurement, field (19) displays a running

 

total of the bit errors.

12. Auto

This field sets the loop delay automatically whenever a bit error test is started.

13/14/15. Loopback These fields select the loopback mode for the mobile station. When one of these fields is selected, the Agilent 8922M/S will attempt to put the mobile station in the described loopback mode. The mobile station must be on a traffic channel to receive the loopback message.

Choices

Off turns loopback mode off.

 

No FE sets loopback mode and turns off frame erasure.

 

FE selects loopback mode and turns on frame erasure.

See Also

Screens: Cell Control

16.Measurement 1-4These fields display a summary of the measurement Type definitions for all measurements 1 through 4. To change these fields, use the Meas Num (6) and Type fields. The Sync Status field displays any errors that occurred while trying to synchronize to the midamble of demodulated data.

Bad Sync appears when the expected midamble bit pattern cannot be found in the demodulated burst.

Possible causes of Bad Syncs are:

Power was too low.

Demodulation trigger too early or too late.

FM (bit) errors found while synchronizing to desired midamble.

RF overload (increase RF Analyzer Amplitude setting).

No Error appears when no synchronization error occurred.

17. Normal

This field allows you to set a normal or a tunable reference.

(Reference)

See Also

Screens: Configure

 

4-13

Page 119
Image 119
Agilent Technologies 8922M, S GSM manual 11/19. Ratio/Count, Auto