Output RF Spectrum, Main View (Option 006 Only)

Output RF Spectrum, Main View (Option 006 Only)

5

3

1

8

2

4

7

6

The Agilent 8922M/S uses a 3-pole synchronously tuned filter to make Output RF Spectrum measurements rather than a 5-pole filter as specified in the GSM recommendations. Refer to Section titled “Output RF Spectrum Measurements Using a 3-Pole Synchronously Tuned Measurement Filter” in Chapter 2.

1.

Amplitude

This is a copy of the RF Analyzer Amplitude field. This is the amplitude to be

 

 

assumed at the selected input port, see RF Analyzer Functions).

 

 

Range

RF IN/OUT: 27.9 to +41.0 dBm. AUX RF IN: 58.0 to +20.0

 

 

 

dBm.

 

 

See Also

Screens: RF Generator/RF Analyzer, (RF Analyzer)

2.

Freq Offset

This sets the frequency offset for the Output RF Spectrum measurement. This field

 

 

is used when not making reference measurements. The offset used is 0.0 kHz when

 

 

Mode is set to Ramp Ref or Mod Ref.

 

 

Range

2.0 MHz to +2.0 MHz.

3.

Sync Status

This field displays any errors that occurred while trying to synchronize to the

 

 

demodulated data.

 

 

 

See Also

Making Measurements: Solving Problems Messages

4-82